
Sterilization in a Flash
Not a quick fix, flash sterilization requires discipline to be
effective
By Charles O. Hancock, RAC
Many OR users consider flash sterilization to be some kind of magic process
that solves a multitude of inventory problems. In fact, use of flash
sterilization requires discipline that few of its users follow. For those
medical devices prepared for use by a Central Sterile Processing Department (CSPD),
it is a given that sterility can be assured and documentation exists to confirm
the status of those devices so prepared. Yet in spite of recommendation to the
contrary1, flash sterilization is routinely carried out in operating
rooms (ORs) every day, often without regard to process performance. Flash
sterilization is the technique used today as an alternative to purchasing
additional instrument sets or to save time when sterile instruments are required
on short notice. Flash sterilization is, in some respects, an orphan process. No
one wants to be responsible for recommending its routine use in the OR.
Nevertheless, flash is widely used in the routine processing of devices within
the OR. Personnel carrying out this flash sterilization process have other
duties which are often considered to be more important than something so mundane
as properly carrying out the complete, disciplined, multi-step process of
decontamination and preparation of the instruments before sterilization. Such
attention to the details of sterilization is left to the personnel in CSPD for
all other goods. Users tend to forget that the flash sterilization cycle is
short in time--often, an assumption is made that no barriers or minimum barriers
to sterilant penetration are present in devices during a flash cycle. That may
not be the case if the users have not paid careful attention to the reprocessing
details required for flash sterilization. The instrument manufacturer may not
have considered the particular requirements of flash sterilization when deeming
the instrument as a reusable medical device. Since reprocessing or
resterilization using flash is not a primary duty of those carrying out the
process, they may be distracted by higher priorities from properly and correctly
carrying out all of the steps necessary for a successful outcome. OR nurses and
technicians are often overburdened with duties.
The exposure times recommended by sterilizer manufacturers are demonstrated
to be effective when used strictly in accordance with their written
recommendations. However, the sterilizer manufacturer assumes no responsibility
for the preparation of the items to be sterilized, nor their proper placement
within a tray or container, nor of the tray's placement within the sterilizing
chamber. Nor do they assume any responsibility for the transport of the items to
the point of use. Sterilizer manufacturers historically have considered flash
sterilization as an act to be done under exceptional conditions. That is, that
flash sterilization will be conducted only when a critical or emergency
situation exists (i.e., the dropped instrument).
This assumption is erroneous. Rather than being the exception, flash
sterilization is the rule in many ORs. (In 1999, one hospital's staff members
said they carried out more than 33,000 flash procedures.4)
Consider that most hospital steam sterilizers are manufactured to meet the
requirements of ANSI/AAMI ST8 Hospital Steam Sterilizers.2 This
American National Standard was developed by the Association for the Advancement
of Medical Instrumentation (AAMI) as a consensus standard reflecting the best
information available and agreed upon by a committee of members representing
users, producers, and general interest categories of expertise. As such, the
standard is accepted by the US Food and Drug Administration (FDA) as a means of
sterilizer manufacturers' compliance to special controls (in this case a
consensus standard) applicable to Class II medical devices. A hospital steam
sterilizer is a Class II medical device. The FDA requires premarket notification
[510(k)] before a hospital steam sterilizer can be introduced to the market.
Part of the information required of the manufacturer for conformance to ST8 is
the documentation of biological testing results specified in the standard. Note
that ANSI/AAMI ST8 currently does not require testing of a "flash" or
unwrapped goods cycle.
Load Specifications
The standard requires and specifies the testing of three types of loads. They
are: the fabric test pack, liquid loads of three one-liter flasks (if
applicable), and a wrapped instrument test pack. Also note that there is
currently no requirement in the standard for a manufacturer to demonstrate a
flash cycle. However, most steam sterilizer manufacturers do make available a
recommendation for a flash cycle based on testing they have done. That testing
defines an unwrapped (or a single wrapper with special instructions) load of
specified size to establish a cycle of sufficient lethality to provide a
sterility assurance level of 10-6. That testing may be for a
simulated small load or even for a single item. Caution should be exercised by
the user, since some manufacturers may not have test data to support the
processing of large tray loads such as those typical in orthopedic instrument
processing. The user should strictly adhere to the sterilizer manufacturer's
written recommendations with respect to that load. If the manufacturer has not
provided such documentation, ask for it. In any event, the user should validate
through a performance qualification that the loads being processed are actually
being subjected to sterilization condition in the configuration presented within
the sterilizer chamber. Any significant changes to that load should be carefully
validated to ensure that the flash process used will sterilize.
The steam sterilizers available and used today for flash sterilization in
hospitals today are designed to be efficient, reliable, and effective when
operated within the manufacturer's recommendations. But, as flash sterilizers
are used and misused, their use may not represent good sterilization practice as
understood by the experts in the field--those responsible for sterilization in
CSPD. There is hope for the user in that AAMI has also provided a recommended
practice that covers flash sterilization in hospitals. ANSI/AAMI ST37 Flash
Sterilization: Steam Sterilization of Patient Care Items for Immediate Use1
was developed to answer this need. This recommended practice concentrates on
what is needed to ensure that flash sterilization is safe and effective. Work
practices are described reflecting the complete process of sterilization as a
guide to those not having a sterilization background. The physical layout of the
facility is considered to provide reference for those planning revision to old
or design of new facilities. There is also discussion regarding procedures to
assure aseptic transfer of sterilized items from the flash sterilizer to the
point of use in the sterile field. Appreciate that a recommended practice such
as ST37 should serve as a guide for those interested in moving toward
performance objectives intended to bring the practice of flash sterilization
more closely into agreement with those practices generally accepted for the
sterilization of reusable medical devices.
When considering any medical/surgical instrument as a reusable device, one
looks to the manufacturer of the instrument for advice and instructions on how
to reprocess that item. That information is especially important when
contemplating an instrument to be a candidate for flash sterilization. The flash
sterilization process is a multi-step process that requires meticulous attention
to detail for success. Failure to decontaminate and clean an instrument properly
can defeat the process. The instrument manufacturer is obligated to provide the
user with at least one method of reprocessing if the device is labeled for
reuse. The method described, however, may not be compatible with your flash
sterilization procedures. The proper disassembly of complex devices is one of
the most difficult manufacturer recommendations to meet for users. Written
procedures are frequently not specific enough to ensure all instruments are
consistently and properly prepared for flash sterilization.
Of particular concern is the sterilization of implantables. Flash
sterilization of implantables is not recommended.3 Orthopedic sets
are typically one of the most frequently flash sterilized sets of instruments.
The items necessitating flash sterilization in those orthopedic sets are items
that are implanted in the patient (i.e., screws).
Careful analysis of items being flash sterilized can potentially identify
alternative solutions that may reduce the number of flash sterilization cycles
being used. For example, if information is collected that accurately reflects
the items being added to a set that then requires flash sterilization, the
reasons for using flash sterilization can be identified. The follow-up action
required is to acquire enough of those items to eliminate the need for flash
cycles.
In addition to providing written documentation of work practices for users to
ensure proper cleaning and decontamination, inspection, and the arrangement and
packaging of instruments in trays to be sterilized, it is recommended that any
OR conducting flash sterilization start tracking by cycle what trays were
processed and why the flash cycle was needed. This analysis will serve to
document the specific needs that must be met to reduce the number of flash
cycles being run. By recording the reason for performing the flash cycle, a
database may be developed that will aid in establishing a rationale for
eliminating the need for the flash cycle in the first place. Often, a small
investment in purchasing additional individual single-use items may result in a
significant reduction in the number of flash cycles being carried out. If those
flash cycles being eliminated are cycles containing implants where biological
monitoring is required, a significant cost reduction can be realized over time.
Anecdotally, the hospital referred to earlier used this technique to reduce the
number of flash cycles from about 33,000 per annum to less than 22,000 in one
year's time.4
Normally OR personnel are too busy to record information concerning their use
of flash sterilization. But they certainly would be happy if they didn't have to
run one of every three flash sterilization they run today. OR managers are well
advised to look at means of eliminating activities such as flash sterilization
since that is the only true means of eliminating labor costs.
Charles O Hancock is the president of Charles O. Hancock Associates, Inc.
in Fairport, NY.
To receive a copy of the ANSI/AAMI ST8 visit www.fda.gov/
For a complete list of references, visit www.infectioncontroltoday.com
For a complete list of references click here
|