Handle With Care
Cleaning and Disinfecting Rigid and Flexible Scopes
By John Roark
Scopes are delicate instruments that must be handled with care
— but cleaned and disinfected with vigilance.
The infection risks associated with both flexible and rigid
scopes is a serious issue. Proper cleaning, disinfection and sterilization of these
instruments must be continually monitored and consistently maintained. Association of periOperative Registered Nurses (AORN)
recommended practices effective Jan. 1, 2003 state:
Personnel should demonstrate competency in the use,
care and processing of endoscopes and related equipment
Equipment should be
inspected in all stages of handling
Equipment should be tested before use and
used according to manufacturers’ instructions
Endoscopes and related equipment
should be cleaned and decontaminated immediately after use, following
manufacturers’ written instructions
Policies and procedures for the processing
of endoscopes, accessories and related equipment should be developed, reviewed
regularly, revised as necessary and readily available in the practice setting.1
Rigid and flexible scopes have different cleaning and
sterilization needs and can pose different threats when it comes to infection
control.
“The rigid scopes are much easier to clean, but pose a
greater risk,” says Lawrence Muscarella, PhD, chief of infection control at
Custom Ultrasonics, Inc. “The flexible scopes are harder to clean, but pose a lower
risk.”
It is important to bear in mind that scopes are made up of
many different components, each requiring special attention. “Each valve, channel and surface of the endoscope in and of
itself is a unique medical instrument,” says Muscarella. “An endoscope is
arguably comprised of seven or eight separate sub-instruments. Each of those
instruments needs to be addressed during reprocessing. Once you understand that
it’s a multi-instrument, you know that there’s a different set of rules that
would apply than if it were a scalpel, which you would consider one instrument.”
Whenever a scope is processed, all channels must be cleaned,
even if they weren’t used during the procedure.
“Problems can occur when someone is not educated,” says
Muscarella. “You can’t train your personnel how to clean the easiest
endoscope and expect them to do the job flawlessly. You need to train that
person on the most complex type of instrument to reprocess, which is probably
the GI colonoscope or GI side-viewing duodenoscope. There has to be attention to
every single channel of the endoscope, whether that channel was or was not used
during the procedure. If you have knowledge of all of the endoscopes and you
understand that different models have different designs and therefore different
needs, the steps will be the same, but the adaptations may be different. You’re going to have a problem if you don’t adapt
properly.”
A basic framework can be followed to help ensure that scopes
are receiving optimal care. Each step is as important as the others, and no step
should be compromised.
Step One: Pre-Cleaning
The first essential step in the cleaning process for both
flexible and rigid scopes is pre-cleaning — wiping down the scope and
immersing it in enzymatic detergent mixed with water immediately after the
procedure to remove any bioburden or debris. Only a detergent that has been
shown by its manufacturer to be effective should be used during endoscope
reprocessing. Do not use a detergent for which data are not available.
Some common mistakes made at this critical juncture, says Lee
Ann Purtell, product manager of rigid scope repair for Mobile Instrument Service
& Repair Inc., is that often, scopes are soaked for too long.
“The general rule of thumb is that no scope should remain
immersed in any solution — even distilled water — for more than 60 minutes.
Follow the manufacturer’s recommended soaking time. When it comes to
enzymatic cleaners, people will think more is better. They’ll put more
enzymatic detergent than is necessary in the mix of water. High concentrations
of detergents can leave surfactants on the glass, which are very hard to remove. They can discolor parts of the scope, and can pit the glass.
Don’t use too high of a mix, and don’t soak too long.”
“A lot of times, scopes will sit in a bin of glutaraldehyde,”
says Rob Purtell, product manager, also with Mobile Instrument Service &
Repair, Inc. “It’s very important not to leave them in longer than the
published time. If a scope is supposed to be in a cold soak for 20 minutes and
you forget about it and leave it in there for several hours, it can begin to eat
away at some of the components. I’ve seen scopes that have been left overnight. It will do a
tremendous amount of damage — follow the published soaking guidelines from the
manufacturers that make that fluid.”
Flexible scopes have an appearance that belies their
fragility, continues Rob. “A lot of times nurses or techs will have the scopes coiled
up almost in knots. The scope naturally flows a certain way if you begin to coil
it up into a sink full of water. There are some weaker points on the tubes that
can be damaged if looped improperly.”
Lee Ann also cautions against soaking scopes beneath other
instruments. “It’s best to keep them separate, and you never want to
put anything heavy on top of the scopes.”
Step Two: Leak Testing
An essential step for flexible scopes is the leak test,
designed to gauge the integrity of the scope, and that no fluids have leaked
into the internal channels. “You need to make sure that the scope is holding
air when you put it through the cleaning process,” says Rob. “If a scope has
a breach that will allow fluid to get inside, it can cause damage to the
internal components. Once the fluid gets in, there’s no way for it to get out —
repair companies have to open up the entire scope and put it in an oven to dry
it out. Inside the scope there are stainless steel components that
will rust, there are also video components and wires that can get corroded.”
Muscarella cites a report of a damaged bronchoscope infecting
a patient. “The hospital failed to perform the leak test, and as a
result, they were transmitting disease from patient to patient,” he says. “Had
they done the leak test, they would have determined that there was damage to the
instrument. It isn’t just fluid invasion that you get — you can get
microbial invasion into any kind of a tear, pit or laceration on the internal
channel. You can’t see down the internal channel — you have no way of
knowing if it was damaged. Leak testing is important not just for the functioning of the
endoscope, but for raising a flag that your instrument may be a potential
carrier of a disease.”
Step Three: Manual Cleaning
Manually brush all channels that can be brushed, says
Muscarella. What you can’t brush, flush with enzymatic detergent (or other
detergent for which efficacy data are available), and soak according to the
detergent’s labeling — paying strict attention to recommended time, dilution
and temperature. “Some of these detergents can cause problems to the scope or
the automated reprocessor,” he says. “You don’t want to immerse the
endoscope in the detergent or the disinfectant for longer than recommended.”
The concept of, ‘if the label says 15 minutes, 30 minutes is twice as good,’
does not apply.
Rigid scopes should be carefully cleaned by hand with a soft
cloth, using care and caution. Lenses can be scratched by an abrasive agent.
Rigid scopes should only be put through instrument washers or used in any
ultrasonic device labeled to be compatible with rigid endoscopes.
Step Four: Rinsing to Remove Detergent and Soil
Lee Ann recommends rinsing scopes two or three times with
distilled water to remove all residual cleaning solutions. “That is a common
mistake,” she says. “People don’t rinse them, and the cleaning
solutions can build up on the lenses. They look hazy or fuzzy, and you can’t
see through them. Use an alcohol wipe to gently remove the surfactants, or
gently use a pencil eraser to erase the film. If it doesn’t come off, it’s
probably pitted the glass, and it will have to be repaired.
“We also suggest at this point prior to sterilization that
the scope be inspected to make sure that all parts are intact, that there’s no
damage to any of the components,” continues Lee Ann. “Use a jeweler’s
magnifying glass to look at the lenses to make sure they are not damaged,
chipped or scratched. Using a scope that has potential damage can weaken the
parts, which can fall out into the patient. It’s important not to be putting
something back into service that is damaged.”
“The cleaning step becomes a real saving grace if the
disinfectant is approaching its minimum effective concentration, or if some
shortcut is taken in the disinfection step,” says Muscarella. “It reduces
the bioburden considerably, so when you go to the next step of disinfecting, you
have much less of a challenge, much less of a formidable problem on your hands.”
Step Five: Disinfecting/Sterilizing
“Sterilizing [rigid scopes] is tricky because there really
is no perfect option currently out there,” says Lee Ann. “Every method
available has its benefits and limitations. ETO gas is probably the least damaging to the scope, and it
will sterilize the instrument. However, people don’t use it because the
turnaround time is prohibitive.”
Both Sterad and STERIS are quick, readily available chemical
methods of sterilization. “With chemical sterilants you can get discoloration,
it can remove any kind of coatings from anodized parts, and depending on the
grade of glass that’s used by the manufacturer, you can get some pitting,”
says Lee Ann, who points out that STERIS is for point-of-use processing only.
“You can STERIS the scopes between cases, but you cannot STERIS at the end of
the day, put it on the shelf, and then use it the next day.”
Steam sterilization and autoclaving present inexpensive
solutions for newer rigid scopes specifically labeled as autoclavable. Older
scopes will be damaged by the heat employed in these processes.
Lee Ann also cautions that immersing hot instruments in cool
water after sterilization will cause thermal shock, and will fracture the
components. For flexible scopes, Muscarella reiterates the importance of
monitoring the temperature, time and concentration of disinfectants.
Step Six: The Second Rinse
Rinsing flexible scopes after disinfecting is an important
means of removing any residual disinfectant. “If you don’t get rid of the
disinfectant that remains, it can cause patient injury,” says Muscarella. The
residues of that disinfectant or detergent can come in contact with the patient,
and cause a reaction by the mucosa of the organ — whether it’s the colon,
the lungs or the esophagus — you’ve got to get the disinfectant off to
protect the patient.”
Muscarella also comments on the ‘rinse water controversy,’
and the importance of monitoring the quality of the rinse after disinfection. “Arguably, the Achilles heel of the entire endoscope
reprocessing protocol is the quality of the rinse water, because it’s the last
thing that’s going to contact that instrument before it’s placed in the
patient,” he says. “If the water itself is contaminated with chemicals, it
can cause a problem. And if it’s contaminated with bacteria, it’s obvious
what it can do.
“I’ve taken the position that you can’t allow any
bacteria in the rinse water because endoscopes and bronchoscopes may sit between
uses, and that bacteria will proliferate. To someone who is immunosuppressed,
introducing just a few bacteria into the lungs can kill them.”
Step Seven: Drying
Thorough drying will break the chain of bacterial transmission
from possibly contaminated water. “You will not get bacteria from the water to
the patient via the endoscope if you dry it,” says Muscarella. “From a
high-level disinfectant and liquid ‘sterilization’ standpoint, it’s
essential to do, because it stops the transmission of any bacteria in the water.
If you’re not monitoring the rinse water, it’s all the more important.”
Step Eight: Storing
Flexible scopes must be stored in a low humidity,
well-ventilated dry area. “You can properly perform all of your cleaning and
disinfection steps, but if you don’t properly store the endoscope, you could
run into a problem,” says Muscarella. “Let’s say you have very, very few
bacteria in your endoscope. If it’s stored in a wet environment, you now allow those
bacteria to double every 20 minutes. In an eight-hour time period, you can go
from an instrument that is completely patient-safe to one that is extremely
hazardous for use simply because you didn’t store it in a dry environment.”
Humidity control is not as much of an issue with the rigid
scopes, but moisture penetration, airborne contaminants and handling that would
cause the package integrity or seals to be broken are factors to be aware of.
Lee Ann stresses the need for carefully handling and storing
scopes, which are more fragile than they appear. “The components inside are
primarily glass,” she says. “Any kind of pressure on the insertion tube —
dropping it, laying something heavy on top, putting any kind of pressure on it
— can break the parts. All the people in central sterile, anyone who handles
the scope really has to pay some special attention and handle them carefully.”
The Bottom Line
Attention to detail, awareness of manufacturer’s guidelines
and keeping an eye out for potential problems are the keys to safer scopes. “I
go to hospitals to train, and I find more and more that people are just not
following the original guidelines by the manufacturer,” says Rob. “When they
buy the scopes, the manufacturer comes in and does the training. The turnover in facilities can be tremendous — new people
come in and the information never gets trickled down the way it should. A lot of
times customers will tell me that the manufacturer last came in six years ago
— that’s the last time they had any formal training on it. The only
consistency I ever see is that I almost always find something. It’s rare to
see a facility that’s really doing things carefully and properly. There’s
usually a breakdown in the process someplace.”
First and foremost is education and training, says Muscarella.
“Make sure that reprocessing personnel have access to the internal schematics
of every single endoscope they have in inventory, ideally with some kind of
chart on the wall, so they can reference each of the models and see what makes
up each scope. Periodic testing and monitoring by supervisors to make sure that
all those scope models are understood is equally important. If the facility does
not ensure its staff is well trained and properly performing each reprocessing
step in accordance with published reprocessing guidelines, it is opening itself
up to a lawsuit. Plain and simple.
“Because GI endoscopy has a very low reported risk of
infection, people get complacent, and corners are cut. The problem with
endoscopes is that they are unlike most other types of instruments. It’s one
reusable instrument that is very complex, and may be used on more than a dozen
patients during the course of a single day. If there is a undetected problem
with any one of the reprocessing steps, there is an exponentially high number of
patients who may have been infected with pathogens. This potential outcome
underscores the importance of periodic training and routine monitoring of each
the facility’s reprocessing steps to ensure staff is in compliance with
published guidelines. Ideally you would like a disposable scenario, but you don’t
have it. Even though the risk appears to be low, people’s guards are down, and
as a result they get a little sloppy. You’re playing with fire when you’re
involved in endoscope reprocessing.”
|