
The Facts:
Wet Packs and Plastic Accessory Cases
By Nancy Chobin, RN, CSPDM; David Furr; and An Nuyttens
Introduction
In the sterile processing community,
there is a perception that plastic reusable instrument cases and trays have a
greater tendency to experience wet pack problems. This article addresses drying
issues with wrapped instrument case and tray sets made out of plastic, metal or
a combination of both.
This paper is the result of a productive and successful
collaboration between the industry (suppliers) and the user (hospitals). The
authors of this paper can highly recommend such collaborative initiatives as
they help both parties to understand better each others’ needs and wants in
the different environments.
Reusable case and tray sets are sterilization containment
devices that need a wrap or have to be enclosed in a rigid container to maintain
sterility during transport. Case and tray sets are mostly customized for
organizing an instrument set. Although less common, generic cases and trays are
also available.
Case and tray sets can be made using all plastic, all metal,
or a combination of both. RADEL® R polyphenylsulfone (PPSU)1 is the principal
high-performance plastic used to manufacture plastic case and tray sets and is
also used for the structural plastic components in hybrid systems (case and tray
sets composed of both metal and plastic). This particular material is used because it is the only
commercially available plastic able to withstand thousands of autoclave cycles
and maintain its impact strength and toughness. Metal case and tray sets are mainly made out of anodized
aluminum, although stainless steel type 304 or 316 is also used when increased
chemical resistance is needed.
Rigid sterilization containers, which use a filter to maintain
a sterile barrier, are not included in this study.
This article publishes the results of drying tests performed
in both an independent laboratory and the central processing department of a
well-known hospital. Several configurations have been evaluated: different
wraps, use of an absorbent towel, customized vs. generic trays, and different
drying times.
This article is a sequel to an earlier article published in
this magazine (“The Facts: Wet Packs and Plastic Accessory Cases,” July 2003 ICT). The first
article showed the results of drying tests performed on a wide variety of
delivery systems, including plastic case and tray sets, metal case and tray
sets, and hybrid systems (combination of metal and plastics). The paper
concluded that when applying good hospital practices, all sets, including
plastic case and tray sets, performed well and did not exhibit wet pack issues.
What is a Wet Pack?
A wet pack problem exists if moisture is exhibited inside the
case and tray upon completion of a sterilization cycle and appropriate cool-down
cycle. Visible moisture may provide a path for microorganisms to enter and
contaminate a wrapped pack.
As is well known in the industry, numerous variables influence
the probability of experiencing wet packs: steam quality, types and sizes of
sterilizers, pack preparation and handling, location of the pack in the
sterilizer, type of the load, weight of the pack, kind of wrap and case and tray
design can be considered as the most important factors.
From a regulatory perspective, the wet pack issue is
complicated by the lack of a generally accepted quantitative definition or test
method for determining when a pack is considered a wet pack. Standards for evaluating sterilization efficacy have been
established by several associations in the United States and in Europe.
For drying requirements, there are no clear guidelines. In
Europe, EN 868, “Packaging materials and systems for medical devices which are
to be sterilized -- Part 8 Re-usable sterilization containers” discusses in
Annex F how to test load drying. However, only textile packs and metal loads are
considered. The European Norm describes the loading of the pack and how to
weight the pack before and after the sterilization and drying cycle. The Norm
allows a weight increase (or water content) as long as it is less than 0.2
percent.
The Association for the Advancement of Medical Instrumentation
(AAMI) is working on a standard, “Containment Devices for Reusable Medical
Device Sterilization” where load dryness is discussed. The draft of the
guideline has been intensely scrutinized and it could take considerable
additional time to finalize the guideline.
DRYING TESTS IN THE LABORATORY: GENERIC
CASE AND TRAY SETS
Procedure
Drying tests were performed at Northview Laboratories, Inc. in
Illinois. Three types of case and tray sets were studied: plastic, metal, and
hybrid (combination of metal and plastic).2 In total, nine sets were tested:
four plastic, three hybrid and two metal. All case and tray sets were generic,
which means that none of the cases and trays were customized for a set of
surgical instruments. None of the trays were organizing trays with specific
locations for instruments. Each set was similarly loaded with miscellaneous
metal surgical tools and instruments to a total pack weight of approximately 20
pounds. The instruments were laid directly on the tray and on the bottom part of
the cases without the use of a mat or towel. After loading, each set was
double-wrapped in heavy duty blue non-woven3 wrap cleared by the Food and Drug
Administration (FDA). The steam sterilizer was a Hotpack. Six packs could be
loaded in the sterilization equipment: three on the top shelf and three on the
bottom shelf of the sterilization rack.
Three different loading patterns were run (see Table 1).

A four-minute, 270-degree F (132-degree C) pre-vacuum
autoclave cycle was run using a 15-minute dry time. Each loading pattern was run
three times. The lab tested nine runs in total. For every run, each case and
tray set was weighed before each autoclave cycle, after the cycle was finished
and again after a 30-minute cool-down period. The sets were then unwrapped and
inspected for the presence of water in the case and tray.
Test Results
None of the packs showed any weight gain when directly weighed
after the autoclaving cycle or after the cool-down period. Furthermore, there
was no wetness on the sterilization wrap following removal from the autoclave.
No water was observed inside the case once the lids were opened and the contents
examined.
Table 2 shows the test results for a single run and is
representative of all nine runs.

click here to view table two
Conclusion
All case and tray sets dried equally with a 15-minute drying
time. No difference in drying behavior could be detected between the plastic,
hybrid and metal cases and trays.
DRYING TESTS IN THE HOSPITAL: GENERIC CASE
AND TRAY SETS
Procedure
In order to confirm the results of the lab study, similar
drying tests with the same generic case and tray sets were performed in the
sterile processing department of Saint Barnabas Medical Center in Livingston,
N.J.
As for the lab study, the case and tray sets were composed of
plastic, metal and metal/plastic (hybrid). Of the nine sets, four were plastic,
three were hybrid and two were metal. All of the loaded sets were weighed and
the weight ranged from 19.9 pounds to 20.2 pounds. None of the trays were organizing trays with specific
locations for the instruments. The various instruments were randomly placed
inside the trays and distributed on both levels. It should be noted that generic
trays may not be the most commonly used in hospital environments. The instruments may have moved around when handled, which
could influence test results.
All of the test cycles were performed in a Getinge USA
prevacuum steam sterilizer Model 3366. The sterilizer was 60 inches in depth.
The sterilizer was installed in 1993 and had 155,528 cycles on it when testing
began. This equipment is not a recent or newer model and was purposely chosen to
reflect common conditions in a clinical environment.
Exposure time in the sterilizer was four minutes at 270
degrees F (132C). Drying times varied depending on the test condition. Each test
condition will be discussed in greater detail in the next section. The packs
were allowed to cool inside the sterilizer for 15 minutes before removing them.
This differs from the lab test, but the cracking of the door is a common
practice in many healthcare facilities. A cool down time of 30 minutes was
applied after removing the packs from the sterilizer. The packs were weighed
before each autoclave cycle, after the cycle was finished and the pack removed
from the sterilizer and again after a 30-minute cool-down time.
The following conditions were tested:
- 20-minute drying time – double lab wrap (blue
non-woven) – no absorbent towels inside the cases and trays
- 30-minute drying time- heavy duty Simul-Wrap4 – no
absorbent towels inside the cases and trays
- 30-minute drying time – heavy duty Simul-Wrap4 –
absorbent towels inside the cases and tray.
Test Results
The first cycle is an exact replica of the lab cycle except
for a 20-minute drying time in the hospital versus a 15-minute drying time in
the lab. Commonly used drying times for similar case and tray sets are minimal
30 to 45 minutes in the hospital. The sets were double- wrapped in the lab wrap
(blue, non-woven) for verification purposes, although this kind of wrapper is
not commonly used for heavy sets in clinical situations. With the lab wrap and a
20- minute drying time, all packs dried and no wet packs were found.
The second cycle used a heavy duty Simul- Wrap®. The drying
time was 30 minutes because the heavy duty wrap presents a greater challenge to
drying cases and trays. No wet packs were observed, however, two packs showed
moisture between the wrapper and the case.
The third cycle tested the drying of the case and tray sets
wrapped in the heavy-duty wrap; however, absorbent surgical towels were placed on the tray and
the bottom of the case. All of the surgical instruments were laid on the towels,
maintaining the same instrument distribution and weight as for the first and
second cycle. After a 30-minute drying time and a 30-minute cool-down time,
the towels were wet but none of the sets had visible moisture inside the set
after drying.
Conclusion
The test results from the independent lab could be replicated.
However, it is important to note that the use of generic trays with a total pack
weight of 20 pounds may not be typical of clinical situations. Because of this,
another series of tests were performed with sets of customized cases and trays.
DRYING TESTS IN THE HOSPITAL: CUSTOMIZED
CASE AND TRAY SETS
Test Procedure
In total, six case and tray sets filled with instrument sets
were tested from three major orthopedic companies: two plastic, two hybrid and
two metal. The weights of the packs varied from 19.1 pounds to 20.4 pounds. The
metal sets were single layer, the plastic sets were either two or three layers
inside the case and the hybrid trays were two layers inside the case.
Six cycles were run in the sterile processing department of
the Saint Barnabas Medical Center, using the same sterilization equipment used
for the generic case and tray sets. Each cycle consisted of a four-minute
exposure time and 30-minute drying time. The door to the sterilizer was opened
at the end of the cycle for 15 minutes, then the autoclave cart was removed. The
items on the cart were placed in a low traffic area (not located near any air conditioning
ducts) and not touched for 30 minutes.
All sets were wrapped with one piece of 45-inch-square
heavy-duty Simul-Wrap. The positions of the sets were rotated in the sterilizer
so that all trays were studied on top or bottom shelves and in the front, middle
and back of the sterilizer. The total number of cycles was six, which means that
the total number of packs going through the sterilization and drying cycle was
36. This high number of packs allows verification of the results.
Test Results
For each of the configurations tested during the six cycles,
all the sets were dry at the end of the sterilization cycle. There was one
exception. One of the plastic case and trays sets containing three layers
exhibited moisture at the outside of the pack. However, this result could not be
replicated during the five following cycles, and the same pack showed dry after
each of these cycles. Although the single incident of wet pack with the set with
three layers could not be repeated, it is understandable that a three-layer
structure (made out either plastic, hybrid or metal) provides more challenge to
dry than a one-layer or two-layer case.
Conclusion
All customized cases dried in the standard hospital
sterilization cycle with a 30-minute drying time. No difference in drying
behavior could be detected between the metal (one layer inside), the plastic
(two or three layers inside) and the hybrid sets (two layers inside).
Summary
A wide variety of delivery systems were tested for wet pack
issues, including all plastic case and tray sets, all metal case and tray sets,
and hybrid systems (combination of metal and plastics). Both generic and
customized case and tray sets were evaluated.
All generic sets contained one case with a lid and one tray
inside. Surgical instruments were placed on the bottom of the case and
in the tray. All packs weighed about 20 pounds. First the drying behavior of
generic containment devices were evaluated in a lab environment. A heavy duty lab wrap (non-woven) was used to wrap the case
and tray. All generic delivery systems were completely dry when applying a
drying time of 15 minutes.
The same generic cases and trays were evaluated in the sterile
processing department of Saint Barnabas Medical Center. The lab results were confirmed.
None of the sets wrapped in the lab wrap showed wet packs with a drying time of
20 minutes. Also, the influence of the quality of the wrap was evaluated. A
heavy-duty wrap demonstrated the need for longer drying times than the double
non-woven lab wrap.
The drying time needed to be extended to 30 minutes for all
the generic sets, due to the greater density of the heavy-duty wrap. Absorbent
towels did not allow improved drying behavior of the sets.
Because generic sets (with a total load of 20 pounds) are not
commonly used in healthcare facilities, customized sets were tested. In total,
six sets of about 20 pounds were tested: two metal, two hybrid and two plastic
cases and trays. All customized packs filled with the specific instrument set
showed similar drying behavior with a drying time of 30 minutes resulting in dry
packs.
In this study, the parameters that proved to have an influence
on drying are the density of the wraps and the design of the set. Generic sets
that are heavily loaded up to 20 pounds do show more challenges to dry than
customized sets. The tests showed that the material choice (metal vs. plastic) does not influence significantly the drying behavior.
Nancy Chobin, RN, CSPDM, is corporate SPD educator at Saint
Barnabas Health Care System in West Orange, N.J. David Furr is with FDC Services
in Fort Wayne, Ind. An Nuyttens is global market manager for Solvay Advanced
|