Table D: Manufacturer's Sterilization Recommendations

Table D: Manufacturer's Sterilization Recommendations
The Recommendations Listed Below Are Found in the Identified Manufacturer's Care and Maintenance Manuals.
INTERPRETING THE DATA:  55/[email protected] C WOULD MEAN 55 MINUTES  OF STERILIZATION AT 132*CENTEGRADE FOLLOWED BY AN 8 MINUTE DRYING CYCLE
NOT RECOMMENDED:  THE MANUFACTURE RECOMMENDS AGAINST THE PROCESS
COMPATIBLE DEVICE:  LISTED IN THE STERRAD "COMPATIBLE DEVICE" DOCUMENT.
NO RECOMMENDATION: 
THE MANUFACTURE MAKES NO CLAIM REGARDING THE PROCESS.
   

UNWRAPPED

WRAPPED

       
INSTRUMENT Max OPERATING AIR PRESSURE GRAVITY PRE-VAC GRAVITY PRE-VAC EtO Gas Plazma Sterris Glutaraldehyde
3M MAXI DRIVER 110psi 55/[email protected] 8/[email protected] 55/[email protected] 8/[email protected]   NO NO NO
3M MINI DRIVER 110psi 55/[email protected] 8/[email protected] 55/[email protected] 8/[email protected]   NO NO NO
ANSPACH BLACK MAX 120psi 20/[email protected] 8/[email protected] 90/[email protected] 10/[email protected] 10:90 [email protected] 3HR / 12HR [email protected] 54*C NO RECOMMENDATION NO NO
ANSPACH BLACK MAX MICRO 120psi 20/[email protected] 8/[email protected] 90/[email protected] 10/[email protected] 10:90 [email protected] 54*C 3HR / 12HR AERATION NO RECOMMENDATION NO NO
CUSA EZCEL 23 & 36 KHZ N/A 30/[email protected]* 4/[email protected] 50/[email protected] 4/[email protected] NO NO NO NO
DYONICS ARTHROSCOPIC SURGICAL SYSTEM N/A 10/0 @132* 4/[email protected]* 30/[email protected]* 4/[email protected]* As Per EtO manufacture's instructions NO RECOMMENDATION NO RECOMMENDATION As Per Glutaraldehyde manufacture's instructions to achieve High Level Disinfection
GORE TUNNELER N/A 10/[email protected]* 4/[email protected]* 15/[email protected] OR 30/[email protected] 3/[email protected] Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL MICRO 100 DRILL 100psi 35/[email protected] or 80/[email protected] 121 4/[email protected] 132 35/[email protected] or 80/[email protected] 121 4/[email protected] 132 Not Recommended NO NO NO
HALL MICRO 100 OSCILLATING SAW 100psi 35/[email protected] or 80/[email protected] 121 4/[email protected] 132 35/[email protected] or 80/[email protected] 121 4/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL MICRO 100 RECIPROCATING SAW 100psi 35min/[email protected]* or 80min/[email protected] 121* 4min/[email protected] 132 35min/[email protected]* or 80min/[email protected] 121* 4min/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL MICRO 100 SAGITTAL SAW 100psi 35min/[email protected]* or 80min/[email protected] 121* 4min/[email protected] 132 35min/[email protected]* or 80min/[email protected] 121* 4min/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL OSTEON DRILL SYSTEM 100psi 35/[email protected] or 60/[email protected] 121 4/[email protected] 132 35/[email protected] or 60/[email protected] 121 4/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL SERIES 4: DRILL/REAMER OSCILLATOR RECIPROCATOR TRAUMA DRILL 100psi 35/[email protected] or 60/[email protected] 121 4/[email protected] 132 35/[email protected] or 60/[email protected] 121 4/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL STERNUM  SAW 100psi 40/[email protected] or 70/[email protected] 4/[email protected] 40/[email protected] or 70/[email protected] 4/[email protected] Not Recommended NO RECOMMENDATION Not Recommended Not Recommended
HALL SURGAIRTOME TWO 100psi 20min/[email protected]* or 40min/[email protected] 121* 4min/[email protected] 132 20min/[email protected]* or 40min/[email protected] 121* 4min/[email protected] 132 Not Recommended NO RECOMMENDATION Not Recommended Not Recommended
HALL ZIMMER AIR DERMATOME   35min/[email protected]* or 60min/[email protected] 121* 4min/[email protected] 132 35min/[email protected] or 60min/[email protected] 121 4min/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
HALL Z-SERTER WIRE DRIVER N/A 15min/[email protected]* or 60min/[email protected] 121* 4min/[email protected] 132 15min/[email protected] or 60min/[email protected] 121 4min/[email protected] 132 Not Recommended NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
MIDAS REX 200psi motor: Spedial Note: Strong cautions against sterilizing dissecting tools except in emergencies. 200psi 15min/[email protected]* or 35min/[email protected] 121* 4min/[email protected] 132 15min/[email protected] or 35min/[email protected] 121 4min/[email protected] 132 18 Hours Minimum Mechanical  Aeration Various Attachments Listed in Sterrad's "Compatible Devices" document NO NO
MIDAS REX 20-120 psi motor: Spedial Note: Strong cautions against sterilizing dissecting tools except in emergencies. 120psi 15min/[email protected]* or 35min/[email protected] 121* 4min/[email protected] 132 15min/[email protected] or 35min/[email protected] 121 4min/[email protected] 132 18 Hours Minimum Mechanical  Aeration Various Attachments Listed in Sterrad's "Compatible Devices" document NO NO
OSTEODICER BY HUNTER 90-120psi 15/[email protected] NO RECOMMENDATION 60/[email protected] NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION
ASTRA: SENSORCAINE-MPF SPINAL INJECTION N/A N/A N/A 121 C(250F) 15min/8dry REMOVE IMMEDIATELY FROM AUTOCLAVE NO RECOMMENDATION NO NO NO NO
SKEETER N/A FOR EMERGENCY ONLY: 3/0 @ 132 Recommended that you use 8minute dry cycle NO RECOMMENDATION 15/8 @ 132 OR 30/[email protected] 4/[email protected] FOLLOW EtO RECOMMENDATIONS; 24 HOUR MINIMUM AERATION "Compatible Device" NO NO
STRYKER 2000 N/A 10/[email protected]* 4/[email protected] 35/[email protected] 4/[email protected]   "Compatible Device" NO NO
STRYKER BONE PLUG HANDPIECE N/A [email protected] [email protected] 35/[email protected] 4/[email protected] 100% [email protected] 120MIN EXP 12HR AERATION NO RECOMMENDATION NO NO
STRYKER TPS DRILL N/A 10/[email protected] 4/[email protected] 35/[email protected] 45/[email protected] 4/[email protected]   NO RECOMMENDATION NO NO
XOMED:KARTUSH STIMULUS DISSECTION INSTRUMENT SET N/A 3/[email protected]   20/8/132 4/8 @ 132 YES:  NO SPECIFIC RECOMMENDATIONS NO RECOMMENDATION NO RECOMMENDATION NO RECOMMENDATION