|Articles|July 30, 2013

Researchers Review Surgical Technology and OR Safety Failures

Surgical technology has led to significant improvements in patient outcomes. However, failures in equipment and technology are implicated in surgical errors and adverse events. Weerakkody, et al. (2013) aim to determine the proportion and characteristics of equipment-related error in the operating room (OR) to further improve quality of care.

A systematic review of the published literature yielded 19362 search results relating to errors and adverse events occurring in the OR, from which 124 quantitative error studies were selected for full-text review and 28 were finally selected.

Median total errors per procedure in independently-observed prospective studies were 15.5, interquartile range (IQR) 2.017.8. Failures of equipment/technology accounted for a median 23.5% (IQR 15.0%34.1%) of total error. The median number of equipment problems per procedure was 0.9 (IQR 0.33.6). From eight studies, subdivision of equipment failures was possible into: equipment availability (37.3%), configuration and settings (43.4%) and direct malfunctioning (33.5%). Observed error rates varied widely with study design and with type of operation: those with a greater burden of technology/equipment tended to show higher equipment-related error rates. Checklists (or similar interventions) reduced equipment error by mean 48.6% (and 60.7% in three studies using specific equipment checklists).

The researchers concluded that equipment-related failures form a substantial proportion of all error occurring in the OR. Those procedures that rely more heavily on technology may bear a higher proportion of equipment-related error. There is clear benefit in the use of preoperative checklist-based systems. We propose the adoption of an equipment check, which may be incorporated into the current WHO checklist.

Reference: Weerakkody RA, et al. Surgical technology and operating-room safety failures: a systematic review of quantitative studies. BMJ Qual Saf doi:10.1136/bmjqs-2012-001778


Related to this article

Water  (Adobe Stock 456102510 by stockphoto-graf)
Understanding current water quality is a starting point for ST108 implementation. John Kimsey, vice president of processing optimization and customer success for STERIS, spoke about his HSPA presentation on assessing readiness, distinguishing system qualification from routine testing, and understanding where water samples should be collected.
AORN26 New Orleans 2026  (Adobe Stock)
Missing instruments and processing delays can strain OR–SPD relationships. At AORN 2026, John Kimsey, vice president of processing optimization and customer success for STERIS, discusses understanding root causes, exploring off-site reprocessing for ambulatory surgical centers, and using hands-on simulation to help surgical teams see sterile processing from a different perspective.
Image of Pathogens flying   (Adobe Stock weerasak)
Behind every landmark in infection prevention, from antimicrobial stewardship guidelines to the COVID-19 town halls that reached flight attendants and teachers, is a generation of physician-scientists who led SHEA through crises, organizational change, and scientific breakthroughs. Infection Control Today traces 30 years of SHEA presidents and the lasting mark they left on the field.