The field often looks to the orthopedic-implant reference value in ISO 19227 — about 500 micrograms of total organic carbon per device — as a starting point, largely because dedicated limits for many device types don't yet exist, a gap worth closing. These limits matter because cleaning is the prerequisite for sterilization: residual organic soil physically shields microorganisms and adds to the burden the sterilant must overcome. In short, you can't reliably sterilize what hasn't first been cleaned.
ICT: Some reusable medical devices are notoriously difficult to clean and reprocess. What design characteristics tend to make a device challenging to reprocess, and what happens during validation when manufacturers discover those challenges?
RD: The features that make a device hard to reprocess are the same ones that make it hard to clean and to sterilize: long, narrow, or dead-ended lumens; hinges, box locks, and mated surfaces that trap soil; rough or porous materials; and internal channels a brush or flush can't reliably reach. In validation, those are exactly where manufacturers place the worst-case challenge—the logic being that if the process works there, it works everywhere on the device.
When a feature fails—soil or bioburden that won't come down to the acceptance limit, or a lumen a low-temperature sterilant such as vaporized hydrogen peroxide penetrates less readily—the manufacturer faces real choices: redesign the feature, add or lengthen cleaning steps, specify different tools, or, if the risk can't be controlled, make the device single-use. FDA's reprocessing guidance expects reprocessability to be designed in, not assumed.
ICT: If an infection preventionist or sterile processing professional wants to look beyond simply following the IFU, what should they understand about the cleaning data and validation evidence behind it? Are there particular details or warning signs they should look for?
RD: The most useful thing to understand is that a cleaning IFU represents a validated worst case: the manufacturer soiled the device more heavily than routine use, cleaned it by those exact steps, and demonstrated that residuals fell below defined limits with a margin. That is why the steps are specific and why deviations matter. If a department wants to look more deeply, reasonable questions to ask a manufacturer include which test soil and endpoints were used, whether the device was tested fully assembled and in its worst-case configuration, and how lumens and mated surfaces were challenged.
The warning signs I would watch for are IFUs that are vague where they should be precise—no defined cleaning agent or concentration, no temperatures or times, no flush volumes for lumens, or a "wipe until visibly clean" instruction for a device with hidden internal spaces. Precise, testable instructions are usually a sign of real validation behind them.
ICT: With vaporized hydrogen peroxide (VHP) receiving greater FDA recognition as a sterilization method alongside ethylene oxide (EO), what should health care facilities understand when comparing the 2 approaches, particularly regarding residuals, aeration, worker safety, device compatibility, and workflow?
RD: Both are low-temperature options for heat- and moisture-sensitive devices, but they behave very differently. Ethylene oxide (EO), governed by ISO 11135, is an excellent penetrant — it reaches long lumens and diffuses through packaging — which is why it remains widely used. Its trade-off is chemistry: EO can leave toxic residuals, including EO itself and, in chloride-containing materials, the byproduct 2-chloroethanol, so devices require aeration and must meet residual limits under ISO 10993-7 before release.
EO is also a recognized human carcinogen with a strict occupational exposure limit, and its facility emissions are the subject of active, evolving regulation. Vaporized hydrogen peroxide (VH2O2), governed by ISO 22441 and recognized by FDA in January 2024 as an Established Category A method, breaks down to water and oxygen. So it leaves little residue, needs minimal aeration, and offers faster turnaround.
Its limits are physical: it penetrates long or narrow lumens less readily, and materials such as cellulosics and nylon absorb it. Neither is universally "safer" —the right choice depends on the device's geometry, materials, and the throughput a facility needs.
ICT: There can sometimes be a disconnect between the people who design and validate reusable devices and the professionals who must reprocess them hundreds or thousands of times in real-world settings. What do you wish manufacturers better understood about sterile processing, and what do you wish sterile processing professionals understood about manufacturers' validation requirements?
RD: From the manufacturing and validation side, what I most wish sterile processing teams understood is why the IFU is written the way it is: every step traces back to something that was tested and proven, so a shortcut that seems harmless—a shorter soak, a different detergent, skipping a lumen flush—quietly breaks the validated process.
The IFU is not bureaucracy; it is the evidence. What I wish manufacturers understood better is the reality on the other side of the bench: the time pressure, staffing, and equipment constraints a department actually works under, and how ambiguous or impractical instructions get in the way of doing the job well.
When an IFU can't reasonably be followed in a real department, that is a design-and-validation problem for the manufacturer to fix, not an adherence failure for the department. The devices that reprocess safely are the ones designed and validated with that real-world workflow in mind.