Table D: Manufacturer's Sterilization Recommendations

October 1, 2000

UNWRAPPED

  

UNWRAPPED

WRAPPED

    
INSTRUMENTMax OPERATING AIR PRESSUREGRAVITYPRE-VACGRAVITYPRE-VACEtOGas PlazmaSterrisGlutaraldehyde
3M MAXI DRIVER110psi55/0@1328/0@13255/20@1328/5@132 NONONO
3M MINI DRIVER110psi55/0@1328/0@13255/20@1328/5@132 NONONO
ANSPACH BLACK MAX120psi20/0@1328/0@13290/20@12110/20@13210:90 MIXTURE@ 3HR / 12HR AERATION@ 54*CNO RECOMMENDATIONNONO
ANSPACH BLACK MAX MICRO120psi20/0@1328/0@13290/20@12110/20@13210:90 MIXTURE@ 54*C 3HR / 12HR AERATIONNO RECOMMENDATIONNONO
CUSA EZCEL 23 & 36 KHZN/A30/0@132*4/0@13250/30@1324/20@132NONONONO
DYONICS ARTHROSCOPIC SURGICAL SYSTEMN/A10/0 @132*4/0@132*30/20@132*4/20@132*As Per EtO manufacture's instructionsNO RECOMMENDATIONNO RECOMMENDATIONAs Per Glutaraldehyde manufacture's instructions to achieve High Level Disinfection
GORE TUNNELERN/A10/O@132*4/0@132*15/20@132 OR 30/20@2503/8@132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL MICRO 100 DRILL100psi35/8@132 or 80/8@ 1214/8@ 13235/8@132 or 80/8@ 1214/8@ 132Not RecommendedNONONO
HALL MICRO 100 OSCILLATING SAW100psi35/8@132 or 80/8@ 1214/8@ 13235/8@132 or 80/8@ 1214/8@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL MICRO 100 RECIPROCATING SAW100psi35min/8dry@132* or 80min/8dry@ 121*4min/8dry@ 13235min/8dry@132* or 80min/8dry@ 121*4min/8dry@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL MICRO 100 SAGITTAL SAW100psi35min/8dry@132* or 80min/8dry@ 121*4min/8dry@ 13235min/8dry@132* or 80min/8dry@ 121*4min/8dry@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL OSTEON DRILL SYSTEM100psi35/8@132 or 60/8@ 1214/8@ 13235/8@132 or 60/8@ 1214/8@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL SERIES 4: DRILL/REAMER OSCILLATOR RECIPROCATOR TRAUMA DRILL100psi35/8@132 or 60/8@ 1214/8@ 13235/8@132 or 60/8@ 1214/8@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL STERNUM  SAW100psi40/8@132 or 70/8@1214/8@13240/8@132 or 70/8@1214/8@132Not RecommendedNO RECOMMENDATIONNot RecommendedNot Recommended
HALL SURGAIRTOME TWO100psi20min/8dry@132* or 40min/8dry@ 121*4min/8dry@ 13220min/8dry@132* or 40min/8dry@ 121*4min/8dry@ 132Not RecommendedNO RECOMMENDATIONNot RecommendedNot Recommended
HALL ZIMMER AIR DERMATOME 35min/8dry@132* or 60min/8dry@ 121*4min/8dry@ 13235min/8dry@132 or 60min/8dry@ 1214min/8dry@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
HALL Z-SERTER WIRE DRIVERN/A15min/8dry@132* or 60min/8dry@ 121*4min/8dry@ 13215min/8dry@132 or 60min/8dry@ 1214min/8dry@ 132Not RecommendedNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
MIDAS REX 200psi motor: Spedial Note: Strong cautions against sterilizing dissecting tools except in emergencies.200psi15min/0dry@132* or 35min/0dry@ 121*4min/0dry@ 13215min/20dry@132 or 35min/20dry@ 1214min/20dry@ 13218 Hours Minimum Mechanical  AerationVarious Attachments Listed in Sterrad's "Compatible Devices" documentNONO
MIDAS REX 20-120 psi motor: Spedial Note: Strong cautions against sterilizing dissecting tools except in emergencies.120psi15min/0dry@132* or 35min/0dry@ 121*4min/0dry@ 13215min/20dry@132 or 35min/20dry@ 1214min/20dry@ 13218 Hours Minimum Mechanical  AerationVarious Attachments Listed in Sterrad's "Compatible Devices" documentNONO
OSTEODICER BY HUNTER90-120psi15/0@132NO RECOMMENDATION60/20@132NO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION
ASTRA: SENSORCAINE-MPF SPINAL INJECTIONN/AN/AN/A121 C(250F) 15min/8dry REMOVE IMMEDIATELY FROM AUTOCLAVENO RECOMMENDATIONNONONONO
SKEETERN/AFOR EMERGENCY ONLY: 3/0 @ 132 Recommended that you use 8minute dry cycleNO RECOMMENDATION15/8 @ 132 OR 30/8@1214/8@132FOLLOW EtO RECOMMENDATIONS; 24 HOUR MINIMUM AERATION"Compatible Device"NONO
STRYKER 2000N/A10/0@132*4/0@13235/8@1324/8@132 "Compatible Device"NONO
STRYKER BONE PLUG HANDPIECEN/A10@1324@13235/8@1324/8@132100% ETO@49-57C 120MIN EXP 12HR AERATIONNO RECOMMENDATIONNONO
STRYKER TPS DRILLN/A10/3@1324/3@13235/8@132 45/8@1214/8@132 NO RECOMMENDATIONNONO
XOMED:KARTUSH STIMULUS DISSECTION INSTRUMENT SETN/A3/0@132 20/8/1324/8 @ 132YES:  NO SPECIFIC RECOMMENDATIONSNO RECOMMENDATIONNO RECOMMENDATIONNO RECOMMENDATION